Correcting heading errors in optically pumped magnetometers through microwave interrogation
Author | |
---|---|
Abstract |
We demonstrate how to measure in situ for heading errors of optically pumped magnetometers (OPMs) in the challenging parameter regime of compact vapor cells with imperfect optical pumping and high buffer gas pressure. For this, we utilize microwave-driven Ramsey and Rabi frequency spectroscopy (FS) to independently characterize scalar heading errors in free-induction-decay (FID) signals. Both of these approaches suppress 5-nT inaccuracies in geomagnetic fields caused by nonlinear Zeeman shifts in FID measurements to below 0.6 nT. |