Correcting heading errors in optically pumped magnetometers through microwave interrogation

Author
Abstract

We demonstrate how to measure in situ for heading errors of optically pumped magnetometers (OPMs) in the challenging parameter regime of compact vapor cells with imperfect optical pumping and high buffer gas pressure. For this, we utilize microwave-driven Ramsey and Rabi frequency spectroscopy (FS) to independently characterize scalar heading errors in free-induction-decay (FID) signals. Both of these approaches suppress 5-nT inaccuracies in geomagnetic fields caused by nonlinear Zeeman shifts in FID measurements to below 0.6 nT.

Year of Publication
2023
Date Published
2024-07
Journal Title
Physical Review Applied
Volume
22
Start Page or Article ID
014005
DOI
URL
Download citation
JILA PI
Associated Institutes
Journal Article
JILA Topics
Publication Status