Alkali-metal optically-pumped magnetometers are prone to inaccuracies arising from the overlap of the average F = I + 1/2 and F = I - 1/2 ground-state Zeeman resonances. We employ density-matrix simulations and experiments to investigate how this hyperfine systematic error varies with spin polarization in a 87Rb free-induction-decay (FID) magnetometer. At low spin polarizations, (P≤0.5), this effect causes single-frequency magnetic-field extraction techniques to exhibit inaccuracies up to approximately 3.5 nT. Density-matrix simulations reveal that this bias can be traced to the relative amplitude and phase between the F = I ± 1/2 hyperfine ground-state manifolds in the FID spin precession signal. We show that this systematic error can be mitigated using either a double-frequency fitting model that accounts for the relative amplitude and phase or synchronous-pulse pumping, that minimizes the F = 1 contribution to the FID signal. Theoretical simulations predict accuracies within 0.5 nT for both techniques across a wide range of spin polarizations, suggesting a sevenfold enhancement over single-frequency extraction methods. Our experiments validate this, showcasing a variation in the extracted field below 1 nT, a 3.5-fold improvement compared to single-frequency extraction methods. Furthermore, the mitigation techniques demonstrate agreement of the extracted magnetic field within 1.5 nT.
BT - Submitted
N2 - Alkali-metal optically-pumped magnetometers are prone to inaccuracies arising from the overlap of the average F = I + 1/2 and F = I - 1/2 ground-state Zeeman resonances. We employ density-matrix simulations and experiments to investigate how this hyperfine systematic error varies with spin polarization in a 87Rb free-induction-decay (FID) magnetometer. At low spin polarizations, (P≤0.5), this effect causes single-frequency magnetic-field extraction techniques to exhibit inaccuracies up to approximately 3.5 nT. Density-matrix simulations reveal that this bias can be traced to the relative amplitude and phase between the F = I ± 1/2 hyperfine ground-state manifolds in the FID spin precession signal. We show that this systematic error can be mitigated using either a double-frequency fitting model that accounts for the relative amplitude and phase or synchronous-pulse pumping, that minimizes the F = 1 contribution to the FID signal. Theoretical simulations predict accuracies within 0.5 nT for both techniques across a wide range of spin polarizations, suggesting a sevenfold enhancement over single-frequency extraction methods. Our experiments validate this, showcasing a variation in the extracted field below 1 nT, a 3.5-fold improvement compared to single-frequency extraction methods. Furthermore, the mitigation techniques demonstrate agreement of the extracted magnetic field within 1.5 nT.
PY - 2024
T2 - Submitted
TI - Investigating the hyperfine systematic error and relative phase in low spin-polarization alkali FID magnetometers
UR - https://arxiv.org/abs/2408.00898
ER -